Power failure
•
Gas failure
•
Apnoea
•
Vte
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
•
During the anaesthesia unit operation place on OFF (
0 ) the main switch (cfr. 6.3.9).
•
The power failure alarm occurs: silence the alarm.
•
The indication " Power failure" is activated.
•
Replace the main switch in ON position ( I ).
•
During anaesthesia unit operation close the medical
gas supply
•
The gas supplì failure alarm occurs: silence the alarm.
•
Restore the medical gas supply.
•
This alarm is activated when the RATE value is equal
to 0.
•
Activate the alarm disconnecting the patient circuit
from patient simulator.
•
After around 20 sec. the alarm is activated, then
silence the alarm.
•
This alarm is activated when the tidal volume value is
higher (lower) than % value of max (minimum) limit
set by operator.
•
Activate the alarm creating a leak in the patient circuit
higher than set alarm limit .
•
After around 20 secs. the alarm is activated, the
silence the alarm.
WARNING! ! Risk of injuries for the patient
The alarms must be activated in the correct times and modes.
Verify the correct activation of the visual and acoustic indications.
Preparation to use
6-49