Annex B: Table 2 - Siare Morpheus Series User Manual

Anesthesia unit
Hide thumbs Also See for Morpheus Series:
Table of Contents

Advertisement

11.4.2 ANNEX B: Table 2

Guidance and manufacturer's declaration – electromagnetic immunity
The Morpheus LT is intended for use in the electromagnetic environment specified below. The customer or
the user of the Morpheus LT should assure that it is used in such an environment.
IMMUNITY test
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
IEC 60601 test level
± 6 Kv contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output lines
±1 kV line(s) toline(s)
±
2 kV line(s) to earth
<5 % UT (>95 % dip in UT)
for 0,5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT)
for 5 s
3 A/m
Compliance level / Virdict
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output lines
±1 kV line(s) to line(s)
±2 kV line(s) to earth
<5 % UT (>95 % dip in UT)
for 0,5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT)
for 5 s
3 A/m
Appendix
Electromagnetic
environment –
guidance
Residential –
Hospital – Other
Residential –
Hospital – Other
Residential –
Hospital – Other
Residential –
Hospital – Other
Residential –
Hospital – Other
11-15

Advertisement

Table of Contents
loading

This manual is also suitable for:

Morpheus lt/mri

Table of Contents