AD.A module
Adjust the O 2 flow.
Verify that the sum of the flows read on all
flowmeters corresponds the foreseen value.
Adjust the concentration of anaesthetic agent
onto the vaporiser.
On the control panel for manual ventilation,
select the type of desired breathing circuit:
AUX: control of fresh gas exit selection
APL : enabling of valves group
The anaesthesia unit is ready to start the phase of patient ventilation
according with the modalities described in the following paragraphs.
Electric control.
The activation allows to enrich the
mixture of gas delivered to the
patient, conveying pure oxygen
(flow of around 35 l/min.) in the
valves group or in the TO and FRO
patient circuit with the aim of
cleaning the circuit in case of
emergency, early awakening, etc.
7-8
MORPHEUS MRI