Siare Morpheus Series User Manual page 197

Anesthesia unit
Hide thumbs Also See for Morpheus Series:
Table of Contents

Advertisement

Maintenance
Frequency
Every 6
months or
1000 working
hours
Every year
Every two
years / when
necessary
10-12
Component
Oxygen sensor
Filters
Patient circuit
Washers – O rings
Anaesthesia unit
Internal battery
To avoid damage to components due to excessive wear,
carry out preventive maintenance and replace parts
following the recommended frequency.
Procedure / Action
Replace.
The working life of the cell depends on the
working environment. If the temperature or the
O2 % is high, the working life of the sensor will
be lower.
Replace.
Sterilize according to the procedure described
in this manual and according to local standards.
Components that cannot be destroyed should
be sterilised before disposal.
Check the performance.
This includes an electrical safety test and
inspection of the ventilator for mechanical
damage and legibility of the labels.
The anaesthesia unit must also be inspected
and checked in general and worn parts must be
replaced, using the appropriate preventive
maintenance kit.
These operations must only be carried out by
qualified technical personnel, according to the
instructions contained in the relative service
and maintenance manual.
Replace.
This operation must only be carried out by
qualified technical personnel, according to the
instructions contained in the relative service
and maintenance manual.
The working life of the battery depends on the
working conditions and environment.
MORPHEUS MRI

Advertisement

Table of Contents
loading

This manual is also suitable for:

Morpheus lt/mri

Table of Contents