S5 Description
4.3.1 Notes
4-8
WARNING! Risk of break of anaesthesia module.
In order to not compromise the correct working or damage the gas
flowmeters, it is necessary to avoid closing of the regulators too
tightly.
The indicated flow value must be read at the upper level of the rim of
the flowmeter indicator when it rotates.
If there is not rotation, call the Siare technical assistance.
Reading point of flowmeter
WARNING! Risk of injury for the patient.
CUT-OFF ALARM. If the anaesthesia module sounds a whistle, it
means that the pressure of the oxygen is too low; immediately taking
action to reset the oxygen pressure.
If the pressure of the main system is not available, use the oxygen
gas cylinder for emergency cases.
Before using the anaesthesia unit on a patient it is necessary to per
form a series of preliminary checks to verify the correct equipment
operation.
Among the preliminary checks to be performed, it is also include the
flowmeter box.
MORPHEUS MRI