Preparation to use
Leakage TEST not overcome
6-38
The losses value of anaesthesia unit is displayed in ml basing on
known and calculated data (flow, pressure and time),
The displayed value must be less than 100 ml.
The values obtained during the LEAK TEST are displayed for the
"compensation of dead space" function.
WARNING – Risk of failure of anaesthesia unit.
This test is not overcome when the pneumatic circuit pressure does
not reach 30 cmH 2 O ( code
E2
100 ml ( code
).
If the preliminary checks phase is not overcome: see cfr. 8 (Alarms)
and cfr. 9 (Troubleshooting).
E1
) or when there is a leakage over
MORPHEUS MRI