7.6

Monitoring

7.6.1 Ventilation parameters monitoring area
Airways
The led bar indicator (with scale from - 10 to + 80 cmH 2 O) displayed in
pressure
real-time the airways pressure during respiratory phase.
PAW
The value shown on the PAW indicates the maximum value of airways
pressure (cmH
VM, Vte,
Area for ventilation parameters displaying, selectable by soft key.
Ti, RATE
VM shows the value of Minute Volume expired by the patient
Vte shows the value of tidal volume expired by the patient
Ti shows the the duration of patient inspiration period
RATE shows the patient breathing rate
When there are differences higher than 20% between set volume (Vt)
and measured value of expired tidal Volume (Vte), it is necessary to
perform the flow sensor calibration.
For the calibration of flow sensor, consult the chapter 5.3.
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
O).
2
AD.A module
7-25