10.2.2 Disinfection and sterilisation

To disinfect the components, dismantle them and place them in a steam disinfection
chamber at 93°C for 10 minutes.
After this first operation and before placing the components in an autoclave, wrap
them in muslin or in a similar material.
Effective sterilization is achieved in an autoclave at 121°C for approx. 15 minutes.
Do not disinfect, sterilize or re-use disposable products.
Disinfect and sterilize every time an infected patient is ventilated.
In normal conditions, disinfect and sterilize according to how often the machine is
used and in any case at least once a month.
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
WARNING! Risk of injury for the patient
Always refer to the instructions provided by the autoclave
manufacturer regarding temperature and time.
Siare recommends:
that the components should be checked every time they are
sterilized and any damaged parts should be replaced.
carrying out a functioning test of the machine whenever parts or
components are replaced.
Maintenance
10-5