8.4.1 Regulations and default values table

Gas supply
High PAW
Low PAW
FiO2%
Apnoea
Vte
Power failure
(main power
supply)
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
The available regulations and the default regulations which can be
selected by the operator are listed here below.
No setting
This entry allows to set the high pressure alarm limit:
10 - 80 cmH 2 O (VAC operative mode)
Pre-set value
5 cmH 2 O
Pre-set value
19%
Pre-set value
It activated with RATE = 0
Settable value
LOFF, L 10% ......... L 90% (step 10% ; by soft key Vt)
Default , L 50%
Pre-set values
100% , 50% , 10%
WARNING! Risk of injury for the user / patient
The anaesthesia units used in the same health environments can
have different preset configurations of alarm limits.
Verify that the preset alarm limits are suitable for the new
patient.
Alarms: AD.A module
8-9