5.2.4 Alarm area
For further information on the operating logic and how setting the
Alarms operating parameters, make reference to the relevant chapter 8.
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
Gas supply failure
High airways pressure
Low airways pressure
Low O 2 concentration
Apnoea
Expired Tidal Volume
Silencing of active acoustic alarm
The battery symbol is combined with orange led indicating that:
•
Battery is charging (with main power supply connected)
•
It indicates that the battery charging level is at 50% (the OFF led
begins to flash orange and you hear a beep) – max operating
autonomy foreseen is around 1.5h.
The battery symbol combined with red led indicates that the battery
charging level is at 25%, the alarm intervenes when the electric
power delivered by the battery is < 10,6 ( the OFF led begins to flash
red and you hear a beep)
AD.A module