Siare Morpheus Series User Manual page 57

Anesthesia unit
Hide thumbs Also See for Morpheus Series:
Table of Contents

Advertisement

S5 Description
Low flows oxygen
rotameter
Low flow nitrous
oxide rotameter
Medical gas supply
Gauges
Alarms
4-4
Scale 0.1 - 1 l/min.
Resolution: 0.05 l/min
Accuracy: ± 10% of read value or: ± 1% of end scale whichever is
the worst case.
Scale: 0.1 - 1 l/min.
Resolution: 0.05 l/min
Accuracy: ± 10% of read value or: ± 1% of end scale whichever is
the worst case.
OXYGEN
Pressure included between 280 kPa and 600 kPa (2,8 - 6 bar)
Max. required flow 90 l/min.
NITROUS OXIDE
Pressure included between 280 kPa and 600 kPa (2,8 - 6 bar)
Max. required flow 15 l/min.
MEDICAL COMPRESSED AIR
Pressure included between 280 kPa and 600 kPa (2,8 - 6 bar)
Max. required flow 90 l/min.
No. 3 on front panel (O
2
Lack or low oxygen pressure with consequent cut-off of Nitrous
Oxide delivery.
- N
O - AIR ), scale 0 - 6 bar
2
MORPHEUS MRI