8.3

Alarms visualization

FAN
Gas Supply
High PAW
Low PAW
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
Not operating fan
No gas supply
High airways pressure
Low airways pressure
It is displayed during the Self Test phase
at ventilator start-up.
It is activated when the gas pressure is
under 2,4 bar
VAC : it can be set through Pmax
parameter
PVC : when it overcomes the pressure
limit of 5 cmH 2 O set through the Pmax
parameter
Fixed value: it is activated when the
pressure is equal or lower than 5 cmH 2 O
Alarms: AD.A module
8-5