5.2.6 Ventilation parameters monitoring area

Airways
pressure
PAW
VM, Vte, Ti,
RATE
FiO 2
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
The led bar indicator (with scale from - 10 to + 80 cmH 2 O) displayed
in real-time the airways pressure.
The value shown on the PAW indicates the maximum value of
airways pressure (cmH
Ventilation parameters displaying are, selectable by soft key.
VM, shows the value of Minute Volume expired by the patient
Vte, shows the value of Tidal Volume Expired by the patient
Ti, shows the duration of the patient inspiration period.
RATE, shows the patient breathing rate
The value shown on the FiO 2 display indicates the percentage value
of O 2 measured which is delivered to the patient.
After the calibration phase (see the relevant procedure by the CAL
button), the display shows the value of 99%.
In case of malfunctioning of the cell and/or its connection interruption
after around 30 seconds the alarm sounds and the relevant led red
lights on. With delivered oxygen percentage lower than 19% the FiO 2
alarm is immediate.