Siare Morpheus Series User Manual page 196

Anesthesia unit
Hide thumbs Also See for Morpheus Series:
Table of Contents

Advertisement

Frequency
Several times
a day /
according to
local practice
and standards
Every day /
when
necessary
Every week /
when
necessary
Every 6
months or
1000 working
hours
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
Component
Patient circuit
Filters
Condensation trap
filter
Flow sensor
Oxygen sensor
Flow sensor
Condensation trap
filter
Anaesthesia unit
Breathing System
Breathing System
Anaesthesia unit
Procedure / Action
Check for any water collection, drain and clean
the tubes when necessary.
Verify their wear.
Check for any water collection, drain and clean
when necessary.
Replace according to procedures and local
directives of reference.
Calibrate according to the procedures
described in this manual.
Replace according to procedures and local
directives of reference.
Check for any water collection, drain and clean
when necessary.
General cleaning and checks.
Dismount the components and clean, then
sterilize in autoclave, disinfect by steam or
chemically.
Dismount the components and clean, then
sterilize in autoclave, disinfect by steam or
chemically.
The anaesthesia unit must be inspected and
checked in general and any worn parts must be
replaced.
Use the appropriate preventive maintenance kit.
This operation must only be carried out by
qualified technical personnel, according to the
instructions contained in the relative service
and maintenance manual.
Maintenance
10-11