Siare Morpheus Series User Manual page 176

Anesthesia unit
Hide thumbs Also See for Morpheus Series:
Table of Contents

Advertisement

Vte
Power failure
(main power
supply)
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
This alarm is activated when the value of expired tidal Volume
is higher (lower) than the value in % of max. (minimum) limit
set by the operator.
It is signalled by an audible and visual alarm. It can be
silenced (suspended) for 20 s.
This alarm indicates the lack or abnormalities on main power
supply.
It is activated after a trigger delay of 0 sec. starting from the
moment when the main power supply is disconnected.
It is signalled by an acoustic and visual alarm. It can be
silenced (suspended) until reaching the 10% of battery
charge.
The power presence LED goes from solid green to flashing
red.
WARNING! Risk of injury for the patient
An auxiliary system of ventilation must be always available for the
patient when the ventilator is for a total life support.
WARNING! Risk of injury for the patient
When this alarm signal is activated there are 10
minutes operation only, after that the machine
stop ventilating.
Alarms: AD.A module
8-11