Preparation to use
The equipment can also work with oxygen supply only, but in this case
the FiO 2 will be adjustable at 99% only.
WARNING! Risk of equipment failure.
In order that the anesthesia machine operates as specified, the inlet
pressures of medical gases must be between 280 kPa and 600 kPa
(2.8 - 6 bar).
Before use ensure that this requirement is met.
WARNING! Risk of power failure.
If all the gas supply hoses are not correctly connected, the anaesthesia
unit will not be available in case of gas supply failure.
Ensure that all supply hoses are connected according with the
indication engraved on the gas inlet bloc and the illustrations in the back
side of the equipment.
After having connected the supply hoses, verify that the anastesia unit
is correctly working.