Maintenance
The instructions for carrying out more detailed tests, for troubleshooting
and for other interventional procedures, information intended for
qualified technical personnel, are contained in the relative chapter.
WARNING! Risk of injury for the user / patient
To ensure the safety of the patient and the operator, the anaesthesia
unit must be inspected and checked when the limit of 1000 working
hours has been reached or, in the event of limited use of the machine,
at least every 6 months.
All maintenance and/or repair operations require perfect knowledge of
the equipment and must therefore only be carried out by highly
qualified personnel, specifically trained and formally authorised by
SIARE.
Inappropriate intervention or unauthorised modifications can
compromise safety and cause danger to the patient.
To avoid the danger of electric shock during maintenance and/or repair
operations, make sure that all power supplies have been disconnected,
disconnect the power supply source (positioning the special danger
signs) and disable all the protection switches of the equipment.
On completion of the maintenance operations, all removed
components should be disposed of according to current waste disposal
regulations.
Components that cannot be destroyed should be sterilised before
disposal.
Follow current regulations for the disposal or recycling of all removed
components.
Before performing the maintenance or repairing works, also in case of
returning the equipment for repairing to manufacturer, it is required to
clean and disinfect the equipment.
10-2
MORPHEUS MRI