Troubleshooting
Apnoea
Flow sensor
calibration
failed
9-6
Check that the sensitivity of the Trigger is appropriate to the patient's
physiological conditions.
Check that the patient circuit is connected correctly to the ventilator
and to the patient.
Contact the nearest Siare Service Centre or a Centre authorised by
Siare.
In this condition, no automatic or spontaneous respiratory is detected (Rate
= 0).
Check that the patient's respiratory parameters are set correctly.
Check that the sensitivity of the Trigger is appropriate to the patient's
physiological conditions.
Check that the mask, endotracheal tube and patient circuit are not in
some way split, disconnected or connected wrongly If this is the case,
eliminate the problem or replace them.
Check that the ventilator functions correctly, checking the airways
pressure trend. If the ventilator functions correctly, check the flow
sensor and the correct connection of its cable.
Check that the patient circuit is connected correctly to the ventilator
and to the patient.
Contact the nearest Siare Service Centre or a Centre authorised by
Siare.
The user can notice directly by monitoring the value of the expired
volume and the peak value of expired flow, if the calibration of flow
sensor was succeeding or not.
Consult the instructions on the user's manual and repeat the flow
sensor calibration verifying eventual errors.
Contact the nearest Siare Service Centre or a Centre authorised by
Siare.
MORPHEUS MRI