Cleaning, Disinfection And Sterilisation Table - Siare Morpheus Series User Manual

Anesthesia unit
Hide thumbs Also See for Morpheus Series:
Table of Contents

Advertisement

10.2.4 Cleaning, disinfection and sterilisation table

Component Procedure
Outer casing Use a moistened disposable cloth with
Screen
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
neutral detergent or a chemical substance
or the like. Use water to remove any
remaining traces of chemical.
The operator may use disinfectants (e.g.
Buraton 10 F, diluted according to the
manufacturer's instructions; VPRO 60C°) to
clean the components.
Disinfectants based on the following
substances can cause damage:
halogen-releasing compounds;
strong organic acids;
oxygen-releasing compounds.
Remove any dust from the surfaces or in
openings using a vacuum cleaner or a soft
cloth.
See above
To avoid damaging the labels and outer surfaces of the
ventilator, use only the chemical substances listed.
Maintenance
Notes
Make sure that no
sprays or liquids
penetrate inside the
equipment and the
connectors.
Do not use cloths or
sponges that could
scratch the surface.