Preparation to use
In the upper side are displayed the measured parameters: accepted
tolerance +/- 10% of set value.
In the lower side are displayed the set parameters.
Before checking the parameters, leave the anaesthesia unit in
operation for at least 10 minutes.
This time allows to the anaesthesia unit to operate at full.
In case of differences over 20% between set volume (Vt) and
measured value of expired tidal Volume (Vte), it is necessary to
perform the calibration of flow sensor (see chap. 5).
WARNING! ! Risk of injuries for the patient
All the pictures and the examples shown in the present chapter have
the mere purpose of being an example and they do not make any
reference to real clinical cases.