User's Manual, DU3310103-MRI / rev. 3 - 02.05.2019
To prevent the risk of fire, keep the lung ventilator and/or the oxygen
tubes of the equipment away from matches, lit cigarettes and
inflammable material, such as anaesthetic gases and/or sources of
heat.
Do not connect the anaesthesia unit to the patient by flexible
connectors, and antistatic or conductive tubes to prevent patient
burnings during the use of high frequency surgical equipment, specially
dangerous with antistatic tubes. The use of flexible connectors,
antistatic or conductive tube is never permitted with MORPHEUS
anaesthesia unit.
Do not use worn and consumed tubes or tubes contaminated by
flammable substances like grease or oil to deliver oxygen; (fabrics, oil
and other fuels can easy ignite and they intensively burn in air with high
concentration of oxygen.
In the event of fire or an unpleasant smell (e.g. a smell of burning), the
lung ventilator should immediately be disconnected from the electrical
power supply and from the battery (if fitted).
When coming into contact with any component of the anaesthesia unit,
the hospital procedures for the handling of infected material should
always be respected.
SIARE is aware that cleaning, sterilisation and disinfection procedures
vary considerably from one health structure to another. SIARE cannot
be held responsible for the efficacy of the cleaning and sterilisation
procedures, nor for the other procedures carried out while the patient is
being treated. As regards cleaning, sterilisation and disinfection of the
product components, it is therefore recommended that the regulations
currently in force in the country where the equipment is installed be
taken into consideration.
The MORPHEUS anaesthesia unit was not designed as a total
monitoring device: some conditions of danger for the patients treated
with vital support equipment will not trigger any alarm.
Before using the anaesthesia unit or any connected component,
carefully check that the equipment is functioning correctly; when
needed, the auto-diagnostic test must be performed as described in the
present User's Manual.
Do not use pointed instruments, such as pencils, screwdrivers or the
like to make selections or settings as they could damage the surface
of the panel.
Check the anaesthesia unit periodically as described in the relative
"Maintenance" chapter and do not use it if it is faulty or malfunctioning.
Replace
any
broken,
contaminated parts immediately, with spare parts supplied by SIARE.
missing,
obviously
worn,
deformed
or
VII