6.3.2 Connection of anesthetic gases scavenging system

Consult the manual relevant to the scavenger manufactured by SIARE.
WARNING! Risk of injury for the patient.
If the scavenging collector of the anaesthesia unit is blocked, in the
breathing system and in patient lungs could create a negative pressure.
Always ensure that the collector is not obstructed and/or blocked.
6-14
21.1
Gas supply connection for active
scavenger
21.2
Morpheus / MRI gas scavenging
connector
21.3
Inlet connector for gas
scavenging from To and Fro
circuit
Note. If the connection is not used, close
the connector on scavenger.
21.4
Outlet connector for gas
scavenging to main system
21.5
Inlet connector for gas
scavenging from Morpheus / MRI
MORPHEUS MRI