Valves group

3.4.2 Connections to valves group

O 2 cell
WARNING. Risk of injury for the user
To avoid risks of electric shocks and/or break of components, during
interventions, verify that the anaesthesia unit has been disconnected
from power supply.
1. Unpack carefully the O 2 cell
2. Pull up the valves group cover.
3. Insert and screw the cell: in the space
marked with the script "O 2 SENSOR".
4. Verify that the electric cable for cell
connection is positioned within the
dedicate space.
Connect the pin to the O2 sensor
Connect the RJ connector on the
dedicated outlet within the valves
group.