I:E
RATE
For further information on the logic and how setting the ventilation
parameters, make reference to the relevant paragraph.
Setting and modification of ventilation parameters
To activate the default settings keep pressed for a few seconds the
membrane key relevant to PRESET settings of required respiratory
parameters.
Modification of Vt parameter
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
Set the relation between inspiratory and expiratory phases.
From 4:1 to 1:4.
Set the ventilator respiratory rate.
From 5 to 90 Bpm (set value in function of max flow value of 80
liters).
Press the key alongside the parameter to be modified.
The parameter value
0500
will start flashing.
(
)
AD.A module
5-15