Preparation to use
6.2.2 Flow sensor assembling
•
Unpack carefully the flow sensor and
the connecting hoses for airways
pressure detection.
•
Insert the flow sensor in the EXP
connector of the valves group.
•
Fix correctly the hoses for pressure
detection from sensor to the
connectors on the anaesthesia unit.
•
Position the hoses for detection of
airways pressure on the patient circuit
supporting arm.
6-6
The flow sensor is single patient type at differential pressure.
Consult the chapter 5 for flow sensor calibration instructions.
WARNING: Danger of injury for the patient.
Check the flow sensor connection .
•
White hose connected to valves group side
•
Blue hose connected to patient group side
MORPHEUS MRI