Annex E: Table 5 - Siare Morpheus Series User Manual

Anesthesia unit
Hide thumbs Also See for Morpheus Series:
Table of Contents

Advertisement

11.4.4 ANNEX E: Table 5

Recommended separation distances between portable and mobile RF communications
equipment and the Morpheus LT
The Morpheus LT is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Morpheus LT can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Morpheus LT as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum
output power of
150 kHz ÷ 80 MHz
transmitter
outside ISM bands
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where
P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note :
1. At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
2. The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to
6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70
MHz.
3. An additional factor of 10/3 has been incorporated into the formulae used in calculating the
recommended separation distance for transmitters in the ISM frequency bands between 150 kHz
and 80 MHz and in the frequency range 80 MHz to 2,5 GHz to decrease the likelihood that
mobile/portable communications equipment could cause interference if it is inadvertently brought
into patient areas.
Note: the values shown in the table refer to the standard levels of the norm, 3V for V
for V
2
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
Separation distance according to frequency of transmitter m
150 kHz ÷ 80 MHz
in ISM bands
3
5 ,
=
d
d
P
V
1
0,12
0,37
1,17
3,69
11,67
80 MHz ÷ 800 MHz
=
12
d
=
P
V
2
0,12
0,38
1,20
3,79
12,00
12,00
Appendix
800 MHz ÷ 2,5 GHz
12
23
=
P
d
E
E
1
0,12
0,23
0,38
0,73
1,20
2,30
3,79
7,27
23,00
and 10V
1
P
1
11-17

Advertisement

Table of Contents
loading

This manual is also suitable for:

Morpheus lt/mri