Siare Morpheus Series User Manual page 84

Anesthesia unit
Hide thumbs Also See for Morpheus Series:
Table of Contents

Advertisement

Setting of Vte alarm limit
WARNING!
After having performed the "Flow Sensor Calibration" it is necessary,
each time, to set the value, in %, of Vte alarm limit.
To set the Vte alarm limits, consult the chapter 8, Alarms.
Verification of spirometry
VAC operative mode:
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
• LOFF, from L 10% to L 90% (step 10% )
• Default , L 50%
Basing on set ventilation parameters check carefully the numerical
data and the parameters calculated by the flow sensor on the
expiratory line.
Vary the set values of the main parameters and check their
correspondence on measured parameters (± 10% tolerance).
AD.A module
5-23

Advertisement

Table of Contents
loading

This manual is also suitable for:

Morpheus lt/mri

Table of Contents