Pmax
In pressure mode it is possible to set the airways pressure limit. In
volumetric modality it is possible to set the limit level of airways high
pressure alarm.
For a better detection of abnormalities on the airways pressure, it is
suggested to set the value of: PAW HIGH is 10 cmH 2 O over peak
pressure reached by the patient.
In such way any obstruction of patient airways or any leak in patient
circuit, it is probably immediately detected by the ventilator.
PEEP
Set the value of airways positive end expiratory pressure
The PEEP is the positive pressure maintained in the patient circuit
during the expiratory phase.
WARNING! Risk of injury to the patient.
The introduction of PEEP during the automatic ventilation requires
caution since it determines the increase of max and mean airways
pressure.
This change requires probably also the modification of alarm limits.
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
Set the pressure limit from: 10 to 80 cmH 2 O.
The value can be set from: OFF, 3 to 30 cmH 2 O.
AD.A module
7-23