Administration Of Fresh Gas In The "To And Fro" System - Siare Morpheus Series User Manual

Anesthesia unit
Hide thumbs Also See for Morpheus Series:
Table of Contents

Advertisement

AD.A module

Administration of fresh gas in the "TO and FRO" system

7.2.2
In order to deliver the fresh gas flow to the TO and FRO system it is necessary that:
the lung ventilator is in MANUAL operative mode;
the TO and FRO patient circuit is connected to the AUX
connector;
that the fresh gas exit selection is activated : enabling of
fresh gas exit connector (AUX) on control panel for
manual ventilation;
that the flowmeter box is correctly set.
CAUTION. Risk of injury to the patient.
In this operative modality there is no monitoring of respiratory vital
parameters: all the patient ventilation phase is entrusted to the User's
sensitivity.
The fresh gas flow must be high to avoid CO 2 accumulation.
7-6