TROUBLESHOOTING
This chapter is a guide for the operator and the technician, providing indications for
eliminating, as quickly as possible, most of the problems that may have caused
malfunctioning or alarm signals.
This chapter describes the possible causes of problems, indicated by alarms that are
activated during normal functioning.
No power
Power supply
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
If the problem persists, carry out a complete check of the
MORPHEUS LT/MRI anaesthesia unit to identify any irregularities.
If the problem cannot be resolved, contact the Siare Service Centre
or a Centre authorised by Siare.
The anaesthesia unit does not switch on
Check that it is connected to the main power supply
Check that the main switch is turned to the I position (ON)
Check the main fuses
Contact the Siare Service Centre or a Centre authorised by Siare.
There is a power supply fault and the anaesthesia unit is operating on
the battery
Check that it is connected to the main power supply
Check that the main switch is turned to the I position (ON)
Check the correct connections of the plug, the fuses and the connector,
and the condition of the cable (if necessary, restore the connections and
replace the cable if it is damaged).
Check that power is present at the relative socket by plugging in another
electrical device. If there is no power, use another socket or check the
overload switch on the electrical panel of the room.
Troubleshooting
9-1