AD.A module
Ventilation parameters displaying
Press the key alongside the ventilation parameters monitoring to display in
sequence other available values.
VM (l)
Vte (ml)
In case of differences higher than 20% between set volume i(Vt) and
measured value of expired tidal volume (Vte), it is necessary to perform
the flow sensor calibration (see following cfr.).
RATE
(bpm)
5-18
Minute volume
Expired Tidal Volume
Respiratory rate
It indicates the value of minute volume expired
by the patient.
The displayed value is in function of the
parameters: Vt and RATE set.
It indicates the value of tidal volume during
the patient expiratory phase.
It indicates the real respiratory rate value
MORPHEUS MRI