AD.A module
I:E
By this parameter it is possible to set the relation between the
inspiratory time and expiratory time. This relation can be set in different
ways upon the respiration type and the mode.
The variation of this parameter modifies the inspiratory flow but not the
tidal volume.
RATE
Set the lung ventilator respiratory rate.
Since the max flow which can be delivered by ventilator is around 80
liters: there is a relation between RATE, Vt and I:E, that prevents the
setting values are not coherent with each other.
For further information on the logic and how setting the ventilation
parameters, make reference to the relevant paragraph.
7-24
Values setting: 4:1, 3:1, 2:1, 1:1, 1:2, 1:3, 1:4.
Values setting from: 5 to 90 Bpm (value which can be set in function
of max flow value of 80 liters).