AD.A module
LEAK TEST
WARNING! - Risk of anaesthesia unit failure and/or injuries for the
patient.
This test verifies that there are no leaks higher than 100 ml in the
anaesthesia unit pneumatic circuits.
WARNING! - Risk of injury for the patient.
Perform the LEAK TEST : consult the "Maintenance" chapter.
STAND-BY operative mode.
Patient circuit correctly mounted on the breathing system.
Close or plug the Y of patient circuit.
Press for a few seconds the TEST membrane key.
The TEST procedure of equipment pneumatic circuit starts.
The value of PAW is going to reach the value of 30 within 15 sec.
At the end of TEST procedure (around 10 sec.) a value in ml, which indicates the
leak of pneumatic circuit, is displayed (see the picture example:
5-10