Preparation to use
Parameters monitoring and spirometry
Basing on set respiratory physiological parameters and on patient characteristics
[patient simulator], the ventilator module can measure and monitor a series of
significative magnitude for clinical evaluation.
Check the correspondence of measured parameters in function of set ones.
The measured values are updated at each breath.
Vary the set parameters values and verify their correspondence on measured
parameters:
PEEP
PAW
VM
Vte
RATE
Ti
FiO 2
6-46
Ensure of the correct operation, checking also that:
the airways pressure increases during the inspiratory phase
the airways pressure limit intervenes (Pmax.: pressometric
operative mode)
the opening of oxygen flowmeter determines the increasing of
concentration on oximeter (FiO 2 %)
at varying of ventilation parameters, the anaesthesia unit
operation corresponds.
the Trigger operation is efficacious.
The anaesthesia unit operates in a coherent way and according
to volume and respiratory rate set values.
MORPHEUS MRI