Preparation to use
Parameters monitoring and spirometry
•
Basing on set respiratory physiological parameters and on patient characteristics
[patient simulator], the ventilator module can measure and monitor a series of
significative magnitude for clinical evaluation.
•
Check the correspondence of measured parameters in function of set ones.
•
The measured values are updated at each breath.
Vary the set parameters values and verify their correspondence on measured
parameters:
•
PEEP
•
PAW
•
VM
•
Vte
•
RATE
•
Ti
•
FiO 2
6-46
Ensure of the correct operation, checking also that:
•
the airways pressure increases during the inspiratory phase
•
the airways pressure limit intervenes (Pmax.: pressometric
operative mode)
•
the opening of oxygen flowmeter determines the increasing of
concentration on oximeter (FiO 2 %)
•
at varying of ventilation parameters, the anaesthesia unit
operation corresponds.
•
the Trigger operation is efficacious.
•
The anaesthesia unit operates in a coherent way and according
to volume and respiratory rate set values.
MORPHEUS MRI