AD.A module
7.4.3 VAC
Respiratory physiological parameters in VAC
Trigger
Vt
RATE
I/E
PEEP
Pmax
The VAC is a volume controlled ventilation (Vt), synchronized with the patient
breaths if the inspiratory trigger is active (Trigger).
This type of ventilation, being completely made by the device, is used when the
patient is not able to breathe autonomously, or, to assure an effective pre-set tidal
volume, it is necessary to ventilate the patient in a fully mechanical way.
7-16
OFF
500 ml
15 bpm
1:2
OFF
30 cmH
O
2
MORPHEUS MRI