Models 707B and 708B Switching Matrix Reference Manual
To specify the type of close, select a Channel Action Type from the box in the upper right before
closing a channel. The options are:
Channel Close: Close the selected channel without affecting the state of any other channels.
Exclusive Slot Close: Close the selected channel and open any closed channels in the same
slot.
Exclusive Close: Close the selected channel and open any closed channels in the instrument
(the only closed channel is the selected channel).
You can open all channels in a slot by clicking Open Slot.
You can open all channels in the instrument by clicking Open All.
For more information on opening and closing channels, see
2-94).
Configure channels from the web interface
To configure channels from the web interface, right-click the channel. The Channel Configuration
dialog box is displayed.
707B-901-01 Rev. A / August 2010
Figure 50: Channel configuration dialog box