Models 707B and 708B Switching Matrix Reference Manual
5. Uninstall other versions of VISA by selecting the appropriate entry from the Control Panel
Add/Remove Programs Wizard list.
6. Restart the Keithley I/O Layer software installation.
7. If the pre-existing version of VISA was supplied by Tektronix or Agilent (as displayed in the
warning message), you may safely reinstall that version of VISA once Keithley I/O Layer software
installation is complete. When you reinstall Tektronix or Agilent VISA, it may prompt you to
preserve the current VISA version, which you should do. This will usually restore the operation of
any dependent applications or drivers.
8. If the pre-existing version of VISA was supplied by a vendor other than Tektronix or Agilent, it is
recommended that you do not reinstall it as this will likely cause the Keithley I/O Layer software to
malfunction.
Installation troubleshooting
If problems occur during installation, it might be helpful to install the components individually. Errors
might appear that will help you resolve the installation issue.
If problems occur during installation:
1. Follow the instructions to uninstall all the KIOL components in
(on page 2-62).
2. Rerun the KIOL installer. Note where the installer unpacks the files (usually in a temporary
folder).
3. Cancel the installer.
4. Go to folder where the files were unzipped.
5. Run the setup.exe for each of the following components in the following order:
IVI Compliance Package (ICP)
Visa Runtime
KIOL
Keithley SCPI Driver
6. Ignore all the other folders.
7. Reboot the computer.
707B-901-01 Rev. A / August 2010
Section 2: General operation
Special installation considerations
2-63