Section 7: Command reference
Example 1
scan.trigger.channel.stimulus =
scan.trigger.EVENT_SCAN_START
Example 2
scan.trigger.channel.stimulus = 0
Also see
scan.trigger.channel.clear()
scan.trigger.channel.set()
Trigger model
scan.trigger.clear()
This function clears the trigger model.
Type
TSP-Link accessible
Function
Yes
Usage
scan.trigger.clear()
Details
This function sets the arm and channel event detectors of the trigger model to the undetected state.
Example
scan.trigger.clear()
Also see
scan.trigger.channel.set()
scan.trigger.channel.stimulus
Trigger model
7-158
(on page 7-156)
(on page 7-156)
(on page 3-1)
Affected by
(on page 7-156)
(on page 7-157)
(on page 3-1)
Models 707B and 708B Switching Matrix Reference Manual
Sets the trigger stimulus of the channel
event detector to scan start event.
Starts the channel action immediately
after the Scan Start Event.
Where saved
Default value
Clears the trigger model.
707B-901-01 Rev. A / August 2010