Tsplink.trigger[N].Clear(); Tsplink.trigger[N].Event_Id - Keithley 708B Reference Manual

Switching matrix
Hide thumbs Also See for 708B:
Table of Contents

Advertisement

Models 707B and 708B Switching Matrix Reference Manual
tsplink.trigger[N].pulsewidth
tsplink.trigger[N].release()
tsplink.trigger[N].stimulus
tsplink.trigger[N].wait()

tsplink.trigger[N].clear()

This function clears the event detector for a trigger.
Type
TSP-Link accessible
Function
Yes
Usage
tsplink.trigger[N].clear()
N
Details
The event detector for a trigger recalls if a trigger event has been detected since the last
tsplink.trigger[N].wait() call. This function clears a trigger event detector, discards the
previous history of the trigger line, and clears the tsplink.trigger[N].overrun attribute.
Example
tsplink.trigger[2].clear()
Also see
tsplink.trigger[N].mode
tsplink.trigger[N].overrun
tsplink.trigger[N].release()
tsplink.trigger[N].stimulus
tsplink.trigger[N].wait()

tsplink.trigger[N].EVENT_ID

This attribute identifies the number that is used for the trigger events.
Type
TSP-Link accessible
Constant
Yes
Usage
eventID = tsplink.trigger[N].EVENT_ID
eventID
N
707B-901-01 Rev. A / August 2010
(on page 7-222)
(on page 7-223)
(on page 7-224)
(on page 7-225)
Affected by
The trigger line (1 to 3)
(on page 7-220)
(on page 7-221)
(on page 7-223)
(on page 7-224)
(on page 7-225)
Affected by
The trigger event number.
The trigger line (1 to 3)
Where saved
Default value
Clears trigger event on synchronization Line 2.
Where saved
Default value
Section 7: Command reference
7-219

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

707b

Table of Contents