Section 2: General operation
Create a configuration script
When you run the create configuration script feature, it automatically generates a user script that is
saved to a script with a name that you define. Create configuration script is available from the front
panel of the instrument, the web interface, and the remote interface.
When you specify the name of the script, be aware that if you specify a name that already exists
(including autoexec), the existing script is overwritten with the new configuration script.
To create a configuration script from the front panel:
1. Press MENU.
2. Select SCRIPT.
3. Select CREATE-CONFIG. The AUTOEXEC ON PWR UP prompt is displayed.
4. Select Yes or No.
5. If AUTOEXEC is set to no, at the name prompt, enter the name of the configuration. The default
name is config01.
6. Press ENTER.
7. The AUTOEXEC message is displayed again. Press EXIT several times to return to the normal
display.
To create a configuration script from the web interface:
1. Open the Unit page.
2. Log in if necessary.
3. Click Create Config Script.
4. To make the configuration script the autoexec script, select Auto-execute on powerup.
5. To assign a name (the script will not be the autoexec script), select Name and enter a name in
the box.
6. Click OK. The configuration script is created.
To create a configuration script from the remote interface:
Send the command:
createconfigscript(name)
Where name is the name you want to assign to the configuration script.
2-108
Models 707B and 708B Switching Matrix Reference Manual
NOTE
Figure 57: Create Config Script dialog box
707B-901-01 Rev. A / August 2010