Section 7: Command reference
Channel patterns are lost when power is cycled.
Channel patterns are stored when you run the createconfigscript() command.
The following restrictions exist when naming a channel pattern:
The name must contain only letters, numbers, or underscores
The name must start with a letter
The name is case sensitive
Examples of valid names:
Channels
Chans
chans
Path1
Path20
path_3
Examples of invalid names:
1path (invalid because it starts with a number)
my chans (invalid because it contains a space)
My,chans (invalid because it contains a comma)
Path1:10 ( invalid because it contains a colon)
An error is generated if:
The name parameter already exists as a label
Any channel is forbidden to close
Insufficient memory exists to create the channel pattern
The parameter string contains slotX (where X equals 1 to 6 for Model 707B, or 1 for Model 708B) or
allslots
The name parameter contains a space character
The pattern name exceeds 19 characters
7-40
Models 707B and 708B Switching Matrix Reference Manual