Models 707B and 708B Switching Matrix Reference Manual
•
scan.scancount
•
scan.trigger.arm.stimulus
•
scan.trigger.channel.stimulus
In addition, the scan list is cleared.
Sending this function only affects the trigger model and scan list settings. To reset the entire system
to factory default settings, use the reset() command.
Example
scan.reset()
Also see
channel.reset()
reset()
(on page 7-140)
scan.bypass
scan.mode
(on page 7-150)
scan.scancount
scan.trigger.arm.stimulus
scan.trigger.channel.stimulus
scan.scancount
This attribute sets the scan count in the trigger model.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
scanCount = scan.scancount
scan.scancount = scanCount
scanCount
Details
During a scan, the instrument iterates through the arm layer of the trigger model the specified number of times.
After performing the specified number of iterations, the instrument returns to an idle state.
707B-901-01 Rev. A / August 2010
(on page 7-43)
(on page 7-145)
(on page 7-151)
(on page 7-154)
(on page 7-157)
Affected by
System reset
Scan reset
The present scan count value (1 to 2,000,000,000)
NOTE
Performs a reset on the trigger model and scan
settings.
Where saved
Default value
Create
1
configuration script
Section 7: Command reference
7-151