Section 7: Command reference
trigger.blender[N].EVENT_ID
This attribute sets the trigger blender event number.
Type
TSP-Link accessible
Constant
Yes
Usage
eventID = trigger.blender[N].EVENT_ID
eventID
N
Details
Set the stimulus of any trigger event detector to the value of this constant to have it respond to trigger events
from this trigger blender.
Also see
None
trigger.blender[N].orenable
This attribute selects whether the blender operates in OR mode or AND mode.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
orenable = trigger.blender[N].orenable
trigger.blender[N].orenable = orenable
orenable
N
Details
This attribute selects whether the blender waits for any one event (the "OR" mode) or waits for all selected
events (the "AND" mode) before signaling an output event.
Also see
trigger.blender[N].reset()
7-202
Affected by
Trigger event number
The blender number (1 or 2)
Affected by
System reset
Trigger blender N reset
The orenable mode:
•
true: OR mode
•
false: AND mode
The trigger blender (1 or 2)
(on page 7-203)
Models 707B and 708B Switching Matrix Reference Manual
Where saved
Default value
Where saved
Default value
Create
false (AND mode)
configuration script
707B-901-01 Rev. A / August 2010