Tspnet.tsp.abortonconnect - Keithley 708B Reference Manual

Switching matrix
Hide thumbs Also See for 708B:
Table of Contents

Advertisement

Models 707B and 708B Switching Matrix Reference Manual
Example
tspnet.tsp.abort(myConnection)
Also see
None

tspnet.tsp.abortonconnect

This attribute contains the setting for abort on connect.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
tspnet.tsp.abortonconnect = value
value = tspnet.tsp.abortonconnect
value
Details
This setting determines if the instrument sends an abort message when it attempts to connect using
tspnet.connect to a TSP-enabled instrument.
When you send the abort command on an interface, it causes any other active interface on that instrument to
close. If you do not issue an abort command (or if tspnet.tsp.abortonconnect is set to 0) and another
interface is active, connecting to a TSP device results in a connection. However, the instrument will not respond
to subsequent reads or executes because control of the instrument is not obtained until an abort command has
been issued. See
Example
tspnet.tsp.abortonconnect = 0
Also see
tspnet.connect()
707B-901-01 Rev. A / August 2010
Affected by
System reset
1 (enable) or 0 (disable)
Communication interfaces
(on page 7-229)
Stops myConnection.
Where saved
Create
configuration script
(on page 4-4).
Configure the instrument so that it does not
send an abort command when connecting to
a TSP-enabled instrument.
Section 7: Command reference
Default value
Enabled
7-237