Digio.trigger[N].Event_Id; Digio.trigger[N].Mode - Keithley 708B Reference Manual

Switching matrix
Hide thumbs Also See for 708B:
Table of Contents

Advertisement

Section 7: Command reference
Example
digio.trigger[2].clear()
Also see
digio.trigger[N].overrun
digio.trigger[N].stimulus
digio.trigger[N].wait()

digio.trigger[N].EVENT_ID

This attribute identifies the trigger event generated by the digital I/O line N.
Type
TSP-Link accessible
Constant
Yes
Usage
eventID = digio.trigger[N].EVENT_ID
eventID
N
Details
To have another trigger object respond to trigger events generated by the trigger line, set the other object's
stimulus attribute to the value of this constant.
Example 1
digio.trigger[5].stimulus = digio.trigger[3].EVENT_ID
Uses a trigger event on digital I/O trigger line 3 to be the stimulus for digital I/O trigger Line 5.
Example 2
scan.trigger.arm.stimulus = digio.trigger[3].EVENT_ID
Uses a trigger event on digital I/O trigger line 3 to be the stimulus for starting a scan.
Also see
None

digio.trigger[N].mode

This attribute describes the mode in which the trigger event detector and the output trigger generator operate on
the given trigger line.
Type
TSP-Link accessible
Attribute (RW)
Yes
7-58
(on page 7-60)
(on page 7-62)
(on page 7-64)
Affected by
The trigger event number
Digital I/O trigger line (1 to 14)
Affected by
System reset
Digital I/O trigger reset
Models 707B and 708B Switching Matrix Reference Manual
Clears the trigger event on I/O line 2.
Where saved
Default value
Where saved
Create configuration script
707B-901-01 Rev. A / August 2010
Default value
digio.TRIG_BYPASS

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

707b

Table of Contents