Trigger.timer[N].Clear(); Trigger.timer[N].Count - Keithley 708B Reference Manual

Switching matrix
Hide thumbs Also See for 708B:
Table of Contents

Advertisement

Models 707B and 708B Switching Matrix Reference Manual
Details
You can set the stimulus of any trigger event detector to the value of this constant to have it respond to
command interface trigger events.
Also see
None

trigger.timer[N].clear()

This function clears the timer event detector and overrun indicator for the specified trigger timer number.
Type
TSP-Link accessible
Function
Yes
Usage
trigger.timer[N].clear()
N
Details
This function sets the timer event detector to the undetected state and resets the overrun indicator.
Also see

trigger.timer[N].count

trigger.timer[N].count
This attribute sets the number of events to generate each time the timer is triggered.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
count = trigger.timer[N].count
trigger.timer[N].count = count
count
N
Details
If count is set to a number greater than 1, the timer automatically starts the next delay at expiration of the
previous delay.
Set count to zero (0) to cause the timer to generate trigger events indefinitely.
707B-901-01 Rev. A / August 2010
Affected by
Trigger timer number to clear (1 to 4)
(on page 7-207)
Affected by
System reset
Trigger timer N reset
Number of times to repeat the trigger
A trigger timer value from 1 to 4
Where saved
Default value
Where saved
Default value
Create
1
configuration script
Section 7: Command reference
7-207