Chapter 2 Command System
When the modulating waveform amplitude reaches the maximum value, the
frequency of the carrier is increased by "frequency offset". When the modulating
waveform amplitude reaches the minimum value, the frequency of the carrier is
decreased by "frequency offset".
Only when FM is selected for the specified function/arbitrary waveform generator
channel, can the frequency offset of FM be set.
Return
The query returns the frequency offset in scientific notation.
Format
Example
:MOD:FM 100
:MOD:FM?
[:SOURce[<n>]]:APPLy?
Syntax
[:SOURce[<n>]]:APPLy?
Description
Queries the current output configuration of the specified function/arbitrary waveform
generator channel.
Parameter
Name
[<n>]
Return
The query returns the current output configuration in "<waveform
Format
name>,<frequency>,<amplitude>,<offset>,<start phase>" format. If no
corresponding parameter is found, DEF is replaced with the specified parameter value.
Example
:APPLy?
[:SOURce[<n>]]:APPLy:NOISe
[:SOURce[<n>]]:APPLy:PULSe
[:SOURce[<n>]]:APPLy:RAMP
[:SOURce[<n>]]:APPLy:SINusoid
[:SOURce[<n>]]:APPLy:SQUare
[:SOURce[<n>]]:APPLy:USER
Syntax
[:SOURce[<n>]]:APPLy:NOISe [<amp>[,<offset>]]
[:SOURce[<n>]]:APPLy:PULSe [<freq>[,<amp>[,<offset>[,<phase>]]]]
[:SOURce[<n>]]:APPLy:RAMP [<freq>[,<amp>[,<offset>[,<phase>]]]]
[:SOURce[<n>]]:APPLy:SINusoid [<freq>[,<amp>[,<offset>[,<phase>]]]]
[:SOURce[<n>]]:APPLy:SQUare [<freq>[,<amp>[,<offset>[,<phase>]]]]
[:SOURce[<n>]]:APPLy:USER [<freq>[,<amp>[,<offset>[,<phase>]]]]
Description
Configures the specified function/arbitrary waveform generator channel to output the
signal with the specified waveforms and parameters.
Parameter
Name
[<n>]
MSO7000/DS7000 Programming Guide
/*Sets the frequency offset of FM of GI to 100 Hz.*/
/*The query returns 1.000000E+2.*/
Type
Range
Discrete
{1|2}
/*The query returns SIN,1000.000000,2.000000,0.500000,90.000000.*/
Type
Range
Discrete
{1|2}
Default
When omitted, by default, the
operations are carried out on GI.
Default
When omitted, by default, the
operations are carried
out on GI.
RIGOL