RIGOL
:TRIGger:PATTern:LEVel
Syntax
:TRIGger:PATTern:LEVel <source>,<level>
:TRIGger:PATTern:LEVel? <source>
Description
Sets or queries the trigger level of the specified channel in Pattern trigger. The unit is
the same as that of the current amplitude.
Parameter
Name
<source>
<level>
Remarks
For VerticalScale, refer to the
the
:CHANnel<n>:OFFSet
Return
The query returns the trigger level in scientific notation.
Format
Example
:TRIGger:PATTern:LEVel CHANnel2,0.16 /*Sets the trigger level of CH2 to 160 mV.*/
:TRIGger:PATTern:LEVel? CHANnel2
:TRIGger:PATTern:SOURce
Syntax
:TRIGger:PATTern:SOURce <source>
:TRIGger:PATTern:SOURce?
Sets or queries the trigger source of Pattern trigger.
Description
Parameter
Name
<source>
Return
The query returns D0, D1, D2, D3, D4, D5, D6, D7, D8, D9, D10, D11, D12, D13, D14,
Format
D15, CHAN1, CHAN2, CHAN3, or CHAN4.
Example
:TRIGger:PATTern:SOURce CHANnel2
:TRIGger:PATTern:SOURce?
2-176
Type
Range
{D0|D1|D2|D3|D4|D5|D6|D7|D8|D9|D10|D
Discrete
11|D12|D13|D14|D15|CHANnel1|CHANnel2
|CHANnel3|CHANnel4}
(-5 × VerticalScale - OFFSet) to
Real
(5 × VerticalScale - OFFSet)
:CHANnel<n>:SCALe
command.
Type
Range
{D0|D1|D2|D3|D4|D5|D6|D7|D8|D9|D10|D
Discrete
11|D12|D13|D14|D15|CHANnel1|CHANnel2
|CHANnel3|CHANnel4}
command. For OFFSet, refer to
/*The query returns 1.600000E-1.*/
/*Sets the trigger source to CH2.*/
/*The query returns CHAN2.*/
MSO7000/DS7000 Programming Guide
Chapter 2 Command System
Default
CHANnel1
0
Default
CHANnel1