Trigger:windows - Rigol DS7034 User Programming Manual

Mso7000 series; ds7000 series; digital oscilloscope
Hide thumbs Also See for DS7034:
Table of Contents

Advertisement

Chapter 2 Command System

:TRIGger:WINDows

Command List:
:TRIGger:WINDows:SOURce
:TRIGger:WINDows:SLOPe
:TRIGger:WINDows:POSition
:TRIGger:WINDows:TIME
:TRIGger:WINDows:ALEVel
:TRIGger:WINDows:BLEVel
:TRIGger:WINDows:SOURce
Syntax
:TRIGger:WINDows:SOURce <source>
:TRIGger:WINDows:SOURce?
Description
Sets or queries the trigger source of Window trigger.
Parameter
Name
<source>
Return
The query returns CHAN1, CHAN2, CHAN3, or CHAN4.
Format
Example
:TRIGger:WINDows:SOURce CHANnel2
:TRIGger:WINDows:SOURce?
:TRIGger:WINDows:SLOPe
Syntax
:TRIGger:WINDows:SLOPe <type>
:TRIGger:WINDows:SLOPe?
Description
Sets or queries the window type of Window trigger.
Parameter
Name
<type>
Remarks
greater than the preset high trigger level.
is smaller than the preset low trigger level.
voltage level meets the preset trigger level.
Return
The query returns POS, NEG, or RFAL.
Format
Example
:TRIGger:WINDows:SLOPe NEGative
:TRIGger:WINDows:SLOPe?
MSO7000/DS7000 Programming Guide
Type
Range
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
Type
Range
Discrete
{POSitive|NEGative|RFALl}
POSitive: triggers on the rising edge of the input signal when the voltage level is
NEGative: triggers on the falling edge of the input signal when the voltage level
RFALl: triggers on either the rising or falling edge of the input signal when the
/*Sets the trigger source to CH2.*/
/*The query returns CHAN2.*/
/*Sets the window type to NeGative.*/
/*The query returns NEG.*/
RIGOL
Default
CHANnel1
Default
POSitive
2-185

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ds7054Ds7024Ds7014Mso7034Mso7024Mso7014 ... Show all