Rigol DS7034 User Programming Manual page 197

Mso7000 series; ds7000 series; digital oscilloscope
Hide thumbs Also See for DS7034:
Table of Contents

Advertisement

Chapter 2 Command System
:TRIGger:RUNT:WHEN
Syntax
:TRIGger:RUNT:WHEN <when>
:TRIGger:RUNT:WHEN?
Description
Sets or queries the qualifier of Runt trigger.
Parameter
Name
<when>
Remarks
NONE: indicates not setting the trigger condition of Runt trigger.
GREater: triggers when the Runt pulse width is greater than the set lower limit of
pulse width.
LESS: triggers when the Runt pulse width is smaller than the set upper limit of
pulse width.
GLESs: triggers when the Runt pulse width is greater than the set lower limit and
smaller than the set upper limit of pulse width.
The lower limit of the pulse width must be smaller than the upper limit.
Return
The query returns NONE, GRE, LESS, or GLES.
Format
Example
:TRIGger:RUNT:WHEN LESS
:TRIGger:RUNT:WHEN?
:TRIGger:RUNT:WUPPer
Syntax
:TRIGger:RUNT:WUPPer <width>
:TRIGger:RUNT:WUPPer?
Description
Sets or queries the upper limit of the pulse width of Runt trigger. The default unit is s.
Parameter
Name
<width>
Remarks
Return
The query returns the upper limit of the pulse width in scientific notation.
Format
Example
:TRIGger:RUNT:WUPPer 0.02
:TRIGger:RUNT:WUPPer?
MSO7000/DS7000 Programming Guide
Type
Range
Discrete
{NONE|GREater|LESS|GLESs}
Type
Range
8.01 ns to 10 s
Real
This command is only available when the qualifier is LESS or GLESs.
When the qualifier is set to "GLESs", the upper limit of the pulse width must be
greater than the lower limit of the pulse width.
/*Sets the qualifier to LESS.*/
/*The query returns LESS.*/
/*Sets the upper limit of the pulse width to 20 ms.*/
/*The query returns 2.000000E-2.*/
RIGOL
Default
NONE
Default
2 μs
2-183